Page top

Industrial Automation

Industrial Automation

Discontinued On Dec. 2015

ZR-RX45

Data Logger

ZR-RX45

200-Channel Expandable Temperature Measurement Logger with Internal 4 GB Flash Memory

* Information in this page is a reference that you created on the basis of information in the product catalog before the end of production, may be different from the current situation, such as goods for / supported standards options / price / features of the product. Before using, please check the compatibility and safety system.

Main Unit

Item ZR-RX45A-[]
Input method All channels isolated input *1
Input terminal shape M3 screw type terminal
Number of analog input channels Standard: 20 ch Max.: 200 ch (When the terminal unit is connected)
Sampling speeds 10 ms (when 1 ch is used) to 1 h
A/D resolution 16 bit
A/D conversion system Delta-sigma
Measure-
ment
ranges
Voltage 20, 50, 100, 200, 500 mV, 1, 2, 5, 10, 20, 50 V, 1-5 V F.S.
Temper-
ature
Thermocouple K, J, E, T, R, S, B, N, W (WRe5-26)
Resistance
temperature
detector
Pt100, JPt100, Pt1000 (IEC751)
Humidity 0% to 100% (Voltage 0 to 1 V scaling conversion) *2
External
input/
output
sections
Digital input Logic input (4 ch) or Pulse input (4 ch) *3 *4
Alarm output 4 ch (Open collector output)
Functions Trigger functions External trigger input (1 ch), Input signal level, time, duration, alarm
Filter functions Off, 2, 5, 10, 20, 40 (moving average)
Calculation function Statistical calculation *5: Average, peak, maximum, minimum, RMS
(maximum of 2 can be set simultaneously)
PCI/F Type LAN (100BASE-TX), USB2.0 (HIGH SPEED)
Function Ethernet
functions
Web server function, FTP server function, FTP client function, NTP client
function, Realtime data transfer to the PC, Unit control from a PC
USB Realtime data transfer to the PC
Memory
devices
Internal Flash memory: approx. 4 GB *6
External USB memory *6
Display 5.7-inch color LCD (VGA: 640 × 480 dots)
Operating environment 0 to 45°C, 5% to 85% (0 to 40°C when operated on batteries, 15 to
35 °C when charging batteries)
Vibration resistance Equivalent to automobile parts Type 1 Category A classification
Power supply AC adaptor; AC100 to 240 V/50 to 60 Hz *7,
DC drive: 8.5 V to 24 V, Battery: DC7.4 V (2 packs)
Power consumption 32 VA (When the AC adaptor is used)
Weight approx. 900 g (Including one terminal unit, excluding batteries and AC
adaptor)
External dimensions 232 × 152 × 50 mm
Accessories User's Manual, Utility disk (CD-ROM), AC Adaptor/AC cable

*1. Resistance temperature detector is non-isolated.
*2. A humidity sensor ZR-XRH1 is necessary.
*3. A logic alarm cable ZR-XRL1 is necessary.
*4. Pulse input: Revolutions mode/Counts mode/Inst. mode switch method
*5. Designate real-time or between-cursors mode (during playback)
*6. 1 file max. 2GB (The capacity will vary depending on the external memory used.)
*7. Be sure to use only the AC cable and AC adaptor provided as accessories.

Measurement Accuracy

Item ZR-RX45A-[]
Voltage ±0.1% of F.S.
Temperature
*1
Thermocouple Measurement Temperature Range (°C) Measurement accuracy
R/S 0 ≤ TS ≤ 100 ±5.2°C
100 < TS ≤ 300 ±3.0°C
R: 300 < TS ≤ 1600 ±(0.05% of rdg + 2.0°C)
S: 300 < TS ≤ 1760 ±(0.05% of rdg + 2.0°C)
B 400 ≤ TS ≤ 600 ±3.5°C
600 < TS ≤ 1820 ±(0.05% of rdg + 2.0°C)
K -200 ≤ TS ≤ -100 ±(0.05% of rdg + 2.0°C)
-100 < TS ≤ 1370 ±(0.05% of rdg + 1.0°C)
E -200 ≤ TS ≤ -100 ±(0.05% of rdg + 2.0°C)
-100 < TS ≤ 800 ±(0.05% of rdg + 1.0°C)
T -200 ≤ TS ≤ -100 ±(0.1% of rdg + 1.5°C)
-100 < TS ≤ 400 ±(0.1% of rdg + 0.5°C)
J -200 ≤ TS ≤ -100 ±2.7°C
-100 < TS ≤ 100 ±1.7°C
100 < TS ≤ 1100 ±(0.05% of rdg + 1.0°C)
N 0 ≤ TS ≤ 1300 ±(0.1% of rdg + 1.0°C)
W 0 ≤ TS ≤ 2000 ±(0.1% of rdg + 1.5°C)

*1. Operating environment 23°C±5°C;
       When 30 minutes or more have elapsed after power was switched on Sampling 1 s/Filter ON (10 times)
       GND connected
       Thermocouple used is T: 0.32 dia., others: 0.65 dia.

Resistance temperature
detector
Measurement temperature
range (°C)
Applied current Measurement accuracy
Pt100 -200 to 850°C (FS = 1050°C) 1 mA ±1.0°C
JPt100 -200 to 500°C (FS = 700°C) 1 mA ±0.8°C
Pt1000 -200 to 500°C (FS = 700°C) 0.2 mA ±0.8°C

Humidity Sensor ZR-XRH1 (Option)

Item Description
Allowable temperature range -25 to 80°C
Allowable humidity range 0% to 100%
Relative humidity measurement accuracy ±3% (5% to 98% at 25°C)
Response time 15 sec (90% response when membrane filter installed)
Sensor output 0 to 1 VDC
External dimensions dia. 14 mm × 80 mm (excluding cable)
Cable length 3 m

PC Software

Item Special PC software ZR-SX10 Standard PC software
Wave Inspire RX (Since Ver.2.4)
(Option Specifications)
Smart Viewer RXW (Since Ver.2.0)
(Standard accessories
Compatible logger ZR-RX25/RX45/RX70/RX20/RX40 ZR-RX25/RX45/RX20/RX40
Compatible operating
system
Windows XP/Windows Vista/Windows 7
CPU Intel-compatible, 1 GHz or faster processor
recommended, Intel Atom CPU not
supported
Pentium 4: 2.0 GHz or faster processor
recommended, Intel Atom CPU not
supported
Memory Windows XP: 512 MB or more Windows
Vista/Windows 7: 1 GB or more
512 MB or more (recommended: 1 GB or
more)
Display 1024 × 768 or higher, screen resolution,
65,535 color (16-bit color) or higher
screen display
1024 × 768 or higher screen resolution
Compatible interface USB, LAN
Standard functions Review saved data, real-time capture of PC data, main unit setup, CSV file conversion
Waveform operation Direct operation of waveform by mouse or
icons, Batch processing of selected
waveform
Change CH scales individually by icons
Waveform display Display graphs in multi-window, X-Y View,
FFT View, Handy function of switching
waveform, Display function of grouping
channel, Scrolling for all directions (up,
down, right, left)
Display in single window, X-Y View between
Cursors (only during replay), Display function
of grouping channel
Configuration function - Channel grouping fuction, Main unit setup
External control of
sampling
Advanced channel grouping function,
Advanced Listview setup function, Main unit
setup
Available (can set and display)
Captured data Binary file (original format): Captured data and the information of graph window are saved.
CSV file: Captured data is saved in comma-separated value format: Binary files can be
converted to CSV files all at once
Others Cursor function, Comment Input function, Excel transfer function

*1: Even when using a PC with adequate specifications, capture errors occur at times because the PC is in bad condition.
       (For example, when other applications are open, or recording medium has no free area.) When you capture data
       using this software, please close all the other applications and save captured data to the hard disk.
*2: Don’t start up other applications while this software is operating, and don’t perform multiple operations. (For
       example, screen saver, virus scan program, copying and moving files, searching files, etc.)